EFFECTIVENESS EVALUATION OF AUTOCLAVES AT THE NATIONAL INSTITUTE FOR CONTROL VACCINE AND BIOLOGICALS

Authors

  • Pham Thi Hang
  • Nguyen Thi Van Quynh
  • Nguyen Ky Anh
  • Le Viet Thanh

DOI:

https://doi.org/10.56086/jcvb.v5i3.227

Keywords:

Wet-steam sterilization, biological indicators, autoclave

Abstract

National Institute for Control of Vaccines and Biologicals (NICVB) has utilized wet-heat sterilization methods for laboratory instruments, solution, and culture media. According to World Health Organization (WHO) recommendations, autoclaves used in sterilization processes require annual revalidation. Therefore, we conducted the study with the titled “Effectiveness Evaluation of Autoclaves at the National Institute for Control of Vaccines and Biologicals to assess the operational performance and sterilization efficacy of autoclaves. The study was performed on two autoclaves: AC 10-TN and AC 11-TN. The results indicated that both autoclaves met operational performance requirements based on physical indicators, including: pressure ≥ 0.1 MPa, temperature ≥ 121°C, holding time ≥ 15 minutes and uniform heat distribution. Sterilization efficacy was also evaluated through biological indicators such as no appearance of microorganisms in both BI cultures and media that ensured sterility, while all positive controls met the required criteria and the risk of false negatives was eliminated. Consequently, both autoclaves were confirmed to be in high quality, reliable, and suitable for use in vaccine and biological product quality control testing

Published

03-10-2025

How to Cite

Pham Thi Hang, Nguyen Thi Van Quynh, Nguyen Ky Anh, & Le Viet Thanh. (2025). EFFECTIVENESS EVALUATION OF AUTOCLAVES AT THE NATIONAL INSTITUTE FOR CONTROL VACCINE AND BIOLOGICALS. JOURNAL OF CONTROL VACCINES AND BIOLOGICALS, 5(3). https://doi.org/10.56086/jcvb.v5i3.227

Issue

Section

Articles

Most read articles by the same author(s)

Similar Articles

1 2 3 > >> 

You may also start an advanced similarity search for this article.