ESTABLISHMENT OF A MOLECULAR PCR METHOD FOR THE DETECTION OF MURINE HEPATITIS VIRUS IN ICR MOUSE MODELS
DOI:
https://doi.org/10.56086/jcvb.v6i2.260Keywords:
Murine hepatitis virus (MHV, RT-PCR, laboratory miceAbstract
Mouse hepatitis virus (MHV) is one of the major viral pathogens requiring surveillance in laboratory mice. Various methods are currently available for MHV detection; among these, reverse transcription polymerase chain reaction (RT-PCR) is well suited to the needs and technical capacity of the Laboratory Animal Department. Using a cross-sectional descriptive design, we compared two primer pairs under different annealing temperatures and MHV concentrations and selected the optimal pair to assess specificity. The results showed that the RT-PCR protocol, employing primers targeting the membrane protein gene, successfully detected MHV at an optimal concentration of 2.8 × 10² TCID₅₀/mL, with an annealing temperature of 60°C.