STUDY ON THE ESTABLISMENT OF A SECONDARY REFERENCE STANDARD FOR THE NEUTRALIZATION ASSAY OF RABIES ANTISERUM POTENCY
DOI:
https://doi.org/10.56086/jcvb.v6i3.276Keywords:
: RAI, rabies immunoglobulin reference standardAbstract
Currently, the only licensed purified equine rabies immunoglobulin in Vietnam is IVACRIG, manufactured by the Institute of Vaccines and Medical Biologicals (IVAC). Potency is tested via the mouse neutralization test (MNT), which requires the international equine rabies antiserum (RAI) standard from NIBSC. Since NIBSC has discontinued this standard, IVAC established and calibrated an in-house secondary standard against the international reference to assign its specific neutralization unit. Methods: A qualified batch of rabies immunoglobulin was adjusted to the required concentration and calibrated against the international RAI standard through ten MNT assays. The geometric mean (GM) and coefficient of variation were calculated using Microsoft Excel to determine the final potency of the secondary standard. Results and Conclusion: A secondary reference standard for the rabies immunoglobulin MNT assay was successfully established with a potency of 39 IU/mL.